KTLD-F
Our company’s electromagnetic flowmeters are manufactured in compliance with the Chinese national standard "JG-1033-2007 Electromagnetic Flowmeters." These instruments serve as inductive devices designed to measure the volumetric flow rate of conductive media. They not only provide real-time on-site monitoring and display but also output a standard current signal, which can be used for recording, regulation, and control purposes, enabling automated detection and control—and even facilitating long-distance signal transmission. This versatile technology is widely applied in industries such as municipal water supply, chemical processing, coal mining, environmental protection, light textiles, metallurgy, and paper manufacturing, where it accurately measures the flow rates of conductive liquids.
Key words:
Classification:
Telephone:
E-mail:
KTLD-F
Overview
Our company’s electromagnetic flowmeters are manufactured in compliance with the Chinese national standard **JG-1033-2007, Electromagnetic Flowmeters**, and are designed as inductive instruments for measuring the volumetric flow rate of conductive media. In addition to providing real-time on-site monitoring and display, these flowmeters can output a standard current signal suitable for recording, regulation, and control purposes, enabling automated detection and control—and even facilitating long-distance signal transmission. They are widely applicable across industries such as water supply, chemical processing, coal mining, environmental protection, light textiles, metallurgy, and paper manufacturing, making them ideal for measuring the flow rates of conductive liquids in these sectors.
How it works
An electromagnetic flowmeter operates based on Faraday's law of induction, which states that when a conductive object moves through a magnetic field, cutting across the magnetic field lines, an induced electromotive force is generated within the conductor. This induced EMF, E, is given by:
E = KBdV
The flow rate Q is:
Q = 3600 × V × S
Where:
K - Instrument Coefficient
B— Magnetic flux density (T)
D—Electrode spacing (m)
V - Average fluid velocity (m/s)
S – Cross-sectional area of the catheter (m²)
When measuring flow, a conductive liquid flows at velocity V through a magnetic field oriented perpendicular to the direction of flow. This motion induces a voltage in the conductive liquid that is directly proportional to the average flow velocity. The induced voltage signal is detected by electrodes that are in direct contact with the liquid. For a given flowmeter, since S, B, and d remain constant, the flow rate is directly proportional to the magnitude of the induced electromotive force E (or the flow velocity V).
Product Features
1. The measuring tube contains no moving parts, making maintenance and management easy; and since there are no flow-obstructing components, there is no pressure loss.
2. The minimum conductivity of the liquid to be measured is ≥30 μS/cm, and when paired with various liner materials, it can be used to measure the flow rate of a wide range of media, including acids, bases, salt solutions, as well as slurries, mineral suspensions, pulp, and more.
3. The measurement of flow is unaffected by changes in the fluid’s density, viscosity, temperature, pressure, or electrical conductivity; instead, the sensor detects a voltage signal.
4. Exhibits a linear relationship with the average flow velocity and offers high measurement accuracy.
5. Proper selection of lining and electrode materials can ensure excellent corrosion resistance and wear resistance.
6. Low-frequency rectangular wave excitation ensures stable and reliable operation, free from interference caused by power frequency and various stray noises present on-site.
7. Independent of fluid direction, it accurately measures flow in both forward and reverse directions.
8. Range ratio of 1:30 (0.5 m/s – 15 m/s), offering a wide full-scale flow velocity range.
9. Chinese character LCD backlit display with online parameter modification, offering simple and convenient operation. 10. Equipped with empty pipe detection and alarm functionality, and capable of adapting to various fluid media.
Complete machine and sensor technical data
Nominal Diameter (mm) |
3/6/10/15/20/25/32/40/50/65/80/100/125/150/200/250/300/350/400/450/500/600/700/800/900/1000/1200/1400/1600/1800/2000.. | ||||
Maximum flow rate |
15m/s | ||||
Accuracy |
DN15-DN600 | Indication accuracy: ±0.5 (flow rate ≥ 1 m/s); ±3 mm/s (flow rate ≤ 1 m/s) | |||
| DN70-DN3000 | Indication accuracy: ±0.5 (for flow rates ≥ 0.8 m/s); ±4 mm/s (for flow rates ≤ 10.8 m/s) | ||||
Nominal pressure |
4.0MPa | 1.6MPa | 1.0MPa | 0.6MPa | 6.3/10MPa |
| DN15-DN50 | DN65-DN250 | DN300-DN1000 | DN1000-DN3000 | Special order | |
Ambient temperature |
Sensor | -25°C to +60°C | |||
| Converter and All-in-One Type | -10°C to +60°C | ||||
Lining material |
Polytetrafluoroethylene, chloroprene rubber, polyurethane, and perfluoroalkoxy alkane (F46) | ||||
Maximum fluid temperature |
All-in-one | 70°C | |||
Separate type |
Perfluoroelastomer lining | 60°C; 120°C (Please specify when ordering) | |||
| Polyurethane lining | 50°C | ||||
| PTFE lining | 100°C; 150°C (Please specify when ordering) | ||||
| Polyfluoroethylene Propylene (F46) | |||||
Electrode material |
Stainless steel, Hastelloy B, titanium, tantalum, platinum/iridium alloy | ||||
Electrode scraper |
Customizable | ||||
Enclosure Protection |
DN15-DN3000 Separable Rubber or Polyurethane-Lined Sensors | IP65 or 1P68 | |||
Spacing (Separate Type) |
Other sensors, integrated flow meters, and split-type transducers | 1P65 | |||
| Converter-to-sensor distances typically do not exceed 30 meters. | |||||
Product Model and Marking

A 1 Installation Type (*Y - Integrated, *F - Split)
B – Nominal Diameter (mm)
c 1 Nominal pressure (MPa) (*06 = 0.6, *10 = 1.0, *16 = 1.6, *40 = 4.0, *XX for special orders)
D— Lining materials (*1—PTFE (F4), *2—CR rubber, *3—polyurethane, *4—PFA copolymer (F46), *5—PFA)
E—Electrode Materials (*G—Stainless Steel, *B—Hastelloy B, *C—Hastelloy C, *I—Titanium, *A—Tantalum, *Y—Platinum-Iridium Alloy, *W—Stainless Steel Coated with Tungsten Carbide)
F- Protection Ratings (*1 - IP65, *2 - IP68 + IP65, meaning sensor IP68 + transmitter IP65; *3 - IP67 + IP65, meaning sensor IP67 + transmitter IP65; *4 - Submersible type, meaning sensor IP68 + transmitter IP68)
G—Explosion-proof marking (*0 - None, *1 - Explosion-proof type)
H—Connection Type (*F-Flange Connection, *S-Screw Connection, *K-Chuck Connection)
I—Power Supply (*1-3.6VDC, *2-24VDC, *3-220VAC, *4-24VDC/220VAC Dual Power Supply)
J-Output Method (*M - Pulse Output, *D - Current Output, *R - RS485, *Q - Full-Function Output, *T - Cooling/Heating Metering)
Coding |
Common diameter |
030 |
3 |
060 |
6 |
100 |
10 |
150 |
15 |
200 |
20 |
250 |
25 |
320 |
32 |
400 |
40 |
500 |
50 |
650 |
65 |
800 |
80 |
101 |
100 |
125 |
125 |
151 |
150 |
201 |
200 |
251 |
250 |
301 |
300 |
351 |
350 |
401 |
400 |
451 |
450 |
501 |
500 |
601 |
600 |
701 |
700 |
801 |
800 |
901 |
900 |
102 |
1000 |
122 |
1200 |
142 |
1400 |
162 |
1600 |
182 |
1800 |
202 |
2000 |
222 |
2200 |
242 |
2400 |
262 |
2600 |
282 |
2800 |
302 |
3000 |
Electromagnetic Flow Measurement Range Selection Table
Caliber (mm) |
Measuring range (m³/h) |
||
10 |
0.02827 - 0.25 |
0.3–1.6 |
2.0–3.3921 |
15 |
0.0636 - 0.6 |
0.8–3.0 |
4.0 - 7.632 |
20 |
0.0636 - 0.6 |
1.2–5.0 |
6.0–13.6 |
25 |
0.176 - 1.6 |
2.0–8.0 |
10-21 |
32 |
0.2895 - 2.5 |
3.0–12 |
16-35 |
40 |
0.4524 - 4.0 |
5.0–20 |
25–45 |
50 |
0.707 - 6.0 |
8.0–40 |
50–85 |
65 |
1.195-10 |
12-60 |
80-143 |
80 |
1.81 - 16 |
20–120 |
160-217 |
100 |
2.83 - 25 |
30–160 |
200-339 |
125 |
4.42 - 40 |
50-250 |
300–530 |
150 |
6.36–60 |
80-400 |
500-763 |
200 |
11.3–100 |
120–600 |
800-1357 |
250 |
17.7 - 160 |
200–800 |
1000–2120 |
300 |
25.45 - 250 |
300–1200 |
1600-3054 |
350 |
34.6 - 300 |
400–1600 |
2000-4157 |
400 |
45.2 - 400 |
500–2000 |
2500-5429 |
450 |
57.3 - 500 |
600–2500 |
3000-6871 |
500 |
70.7 - 600 |
800–3000 |
4000-84823 |
600 |
102-800 |
1000–4000 |
5000 - 12216 |
700 |
139-1200 |
1600–5000 |
6000-16620 |
800 |
181-1600 |
2000–6000 |
8000-21720 |
900 |
229-1600 |
2000–8000 |
10,000 - 27,480 |
1000 |
283-2000 |
2500–10,000 |
12,000 - 33,924 |
1200 |
407-2500 |
3000–12,000 |
16,000 - 48,833 |
1400 |
554-3000 |
4000–16,000 |
20,000 - 66,468 |
1600 |
723-4000 |
5,000–20,000 |
27,000 - 86,815 |
Factory display
Plant display 7
Plant display 8
Plant display 6
Plant display 5
Plant Display 4
Plant Display 9
Plant Exhibition 3
Plant display 2
Factory display
Honor Show
Exhibition style
Previous Page
KTLD-TF
Next Page
Product Consulting
We will contact you within one working day. Please pay attention to your email.